This amplifier can exhibit power performance for broadband operation (DC-100 kHz). AM30AZ is a general-purpose amplifier with an extremely low drift, although it achieves high gain so that voltage can be amplified up to 10000 times. Equipped with a wide range of calibration and a low-pass filter, it can be applied to various experiments and measurements. AM30AZ is also equipped with the electronic auto zero function that allows a burdensome zero point adjustment to be made by one-touch operation. It is convenient for measurement with multi-channels.
Specifications
| Compatible input resistance | 60Ω - 2kΩ |
|---|---|
| Sensor excitation power supply | Voltage; 2.5, 5, 10, 15, 20V DC ±0.2% Current; max 50mA |
| Equilibrium adjustment range | ±1.5mV/V (excitation voltage 10V) |
| Input voltage range | 0 - ±100mV |
| Input resistance | 10MΩ or more (direct-coupled equilibrium differential) |
| Voltage amplification | Max. 10000 times |
| Sensitivity adjustment | - Coarse adjustment: 0, 1/20, 1/10, 1/5, 1/2, 1/1…Error±0.2% - Fine adjustment: 1 to 1/6 of each range variable |
| Zero drift | ±0.5μV/°C (by input conversion) (after a lapse of 30 min. after power ON) |
| Sensitivity variation | Within ±50 ppm/°C of measured value (after a lapse of 30 min. after power ON) |
| Noise | Full band: Input conversion within 30μVp-p(RTI) 1 kHz band: Input conversion within 2.5μVp-p(RTI) 10 Hz: Input conversion within 0.5μVp-p(RTI) |
| Non-linearity | Within ±0.01% FS |
| Output voltage | 0 - ±10V ±5 mA, Output resistance 0.2Ω (BNC connector output) 0 - ±10V ±30 mA, Output resistance 300Ω (binding post output) |
| Response characteristic | DC -100 kHz / -3dB |
| High-pass filter (AC mode) | 0.5 Hz (-3dB) ±15% -6dB/oct |
| Low-pass filter | 1, 10, 100, 1k, 10 kHz (-3dB) ±15% and PASS -12dB/oct, Bessel type |
| Calibration voltage | ±0.125, 0.25, 0.5, 1.25, 2.5mV/V ±0.2%, superimposed on input signal |
| External control | Electronic auto zero function - Auto zero range: ±1V (±10V, ±5V, ±2V Please specify when ordering.) - Auto zero command: ON and OFF, panel face switch and external signal (negative pulse with a pulse width of 50ms or more) |
| Operating conditions | ・ Ambient temperature: 0 - +50°C ・ Ambient humidity: within 0-85% RH (non-condensing) |
| Required power supply | 100V AC ±10%, 50/60 Hz, approx. 15VA |
| External dimensions | 49.5 (W) x 138 (H) x 310 (D) [mm] (panel thickness included, but dimensions of protruding parts not included) |
| Weight | Approx. 1.9 kg |
| AUTO ZERO PERFORMANCE | |
| Auto zero operation range | ±10% FS |
|---|---|
| Internal basic resolution | 12bit (±1/2000) |
| Auto zero command | Via front panel ON/OFF switch or external control signal (negative pulse with a pulse width of 50mS or more) |
The electronic auto zero function allows instant zero adjustment with the push switch on the front panel or external control signal.
Optional Accessories
| Model | Description |
| CA3Y31 | (6-core) input cable with PRC connector at one end (3 m) |
|---|---|
| CA451 | (6-core) input cable with PRC connector at both ends (5 m) *custom manufactured product |
| CA7 | Coaxial cable with BNC |
| CA8 | Cable with alligator clip at one end |