F320
Compact DIN 96 x 48 mm Size Digital Indicator
Product Outline
F320 is most ideal in controlling production line quality.
F320 is a digital indicator with DIN 96 x 48 mm size for strain gauge type sensors. It is most ideal in controlling load pressure such as in pressing, compression and caulking work; as a torque controller of a nut runner or rotary devices and in production line quality control. It is built-in with the RS-485 (Modbus-RTU protocol) that allows easy construction of a control system that is integrated with a PLC or programmable display device.
Specifications
ANALOG SECTION|DISPLAY SECTION|SETTING SECTION|EXTERNAL INPUT/OUTPUT SIGNAL|INTERFACE|OPTION|GENERAL SPECIFICATIONS|Accessories
| ANALOG SECTION | |
| Sensor excitation voltage | 10V or 2.5V DC ±10% (2.5V default setting); Output current: within 30mA |
|---|---|
| Signal input range | -3.0 - +3.0mV/V |
| Equiv. input cal. range | +0.5 - +3.0mV/V |
| Equiv. input cal. error | within 0.1% FS (at 0.5mV/V input) |
| Zero adjustment range | -2.0 - +2.0mV/V |
| Gain adjustment range | +0.5 - +3.0mA / V |
| Min. input sensitivity | 1μV / count (1/10000 guaranteed at 1mV/V input) |
| Accuracy |
|
| A/D converter |
|
| Analog filter | 3, 10, 30, 100, 300, 1 kHz, low-pass filter (-6dB/oct.) |
| Peak hold function | Digital peak hold at 2000 times/sec |
| DISPLAY SECTION | |
| Display unit |
|
|---|---|
| Display items |
|
| SETTING SECTION | |
| Key switch | FNC (+ / -), GAIN, CAL, ZERO, HOLD |
|---|---|
| Setting items | Calibration: Zero / Span calibration (Actual load / Equivalent input calibration) HI limit, LO limit, HI/LO limit comparison mode, hysteresis, digital offset, near zero Digital filter, analog filter, motion detect (time), motion detect (range). Zero tracking (time), zero tracking (range), hold mode Set value LOCK, calibration value LOCK, ZERO key ON/OFF, HOLD key ON/OFF, min. scale, display frequency, decimal place, excitation voltage RS-485 communication condition, RS-485 ID number, RS-485 transmission display time, BCD data update rate, D/A converter zero setting, D/A converter full scale setting, D/A output mode, password |
| EXTERNAL INPUT/OUTPUT SIGNAL | |
| HI limit relay output, LO limit relay output, hold signal input, digital zero signal input | |
| INTERFACE | |
| RS-485 communication interface (Modbus-RTU communication protocol) | |
| OPTION | |
| BCD parallel data output interface (sink type) BCD parallel data output interface (source type) D/A converter (voltage output) D/A converter (current output) |
|
| GENERAL SPECIFICATIONS | |
| Power supply voltage | 100V - 240V AC +10% - 15% (free power supply), 50/60 Hz, 12 - 24V DC (±15%) (Please specify when ordering.) |
|---|---|
| Power consumption | 15W max |
| Inrush current (Typ): AC specifications | 15A, 2msec: 100V AC average load condition (cold start at room temperature) 30A, 2msec: 200V AC average load condition (cold start at room temperature) |
| Inrush current (Typ): DC specifications | 4A, 25msec: 12V DC average load condition (cold start at room temperature) 3A, 20msec: 24V DC average load condition (cold start at room temperature) |
| Operating conditions | Temperature: Working temperature range -10 - +40°C; Storage temperature range -40 - +80°C Humidity: 85% RH or less (non-condensing) |
| External dimensions | 96 (W) x 48 (H) x 127.3 (D) [mm] (protruding parts not included) |
| Panel cutout dimensions | 92 x 45 +1 -0 [mm] |
| Weight | 0.7 kg |
| Accessories | |
| ・AC power supply cord* 1 ・Terminator 1 ・3P-2P conversion adapter* 1 ・BCD output connector (with BCD option) 1 ・Ferrite core for sensor cable 1 ・Ferrite core for AC power supply cord* 1 ・F320 Operation Manual 1 *Included only for AC specifications |
|
Option
| Model | Description |
| BCO | BCD parallel data output interface (Sink type) |
|---|---|
| BSC | BCD parallel data output interface (Source type) |
| DAV | D/A converter voltage output |
| DAI | D/A converter current output |
Optional Accessories
| Model | Description |
| CAAC3P-B3 | AC Supply cord 3m |
|---|---|
| CAAC3P-CEE7/7-B2 | AC Supply cord (Voltage resistance:250V) 2m |
| CN3P-2P | 3P-2P converter plug for AC input cord |
| CN51 | BCD output connector |
| TSU01 | Lightning surge unit |
| TSU03 | Lightning surge unit, DC specification |
Download
| Product Catalogue | Operation Manuals | External dimentions | Support Tools | Software |
|
|

- Download an operation manual, product appearance diagram, support tool or software after user registration.