LC240

Loadcell Converter
 

Product outline

LC240 is a loadcell converter with the large capacity that allows up to 4 loadcells to be driven.

LC240 is a low-drift loadcell converter designed for loadcells.
With a U300DC preamplifier with a proven track record in stability adopted for the head amplifier, it realizes low drift and low noise features.
The bridge excitation power supply can be adjusted to 6 different levels besides the large capacity that allows up to 4 loadcells to be driven in parallel. LC240 is capable of loadcell converter that is standard equipped with the calibration function and current output.

Loadcell Converter LC240
 
 

Specifications

Excitation voltage 10V DC ±0.5%, Output current: within 120mA…..Standard specification
(Up to 4 loadcells at 350Ω can be parallel-connectable.)
2.5, 2.9, 4, 5, 6,6V DC ±0.6%
(Switching over via DIP switch)
Output current: within 60mA
Zero adjustment range 0 - approx. 2mV/V
  • Coarse adjustment:
  • 16-step rotary switch
  •  
  • Fine adjustment:
  • 20 turn trimmer potentiometer
Gain adjustment range 0.45mV/V - 3.0mV/V 0.225mV/V - 1.5 mV/V (Range can be changed by selecting an internal switch)
  • Coarse adjustment:
  • 16-step rotary switch
  •  
  • Fine adjustment:
  • 20 turn trimmer potentiometer
Accuracy
  • Non-linearity:
  • within 0.01%/FS
  •  
  • Zero drift:
  • within 0.25μV/°C CRTI
  •  
  • Gain drift:
  • within 50ppm/°C
  •  
  • Noise:
  • within 0.2μVp-p RTI (1 Hz)
Output signal Voltage output: 0 -±10V, Load resistance: 2kΩ or more, Output resistance: 0.2Ω or less
Current output: 4 - 20mA, Load resistance: 500Ω or less
Low-pass filter Low-pass filter (-6dB/oct.)
Approx. 1 Hz / -3dB….Standard specification
Calibration value 1 mV/V ±0.2%....Standard specification
Power supply voltage 100V AC ±10%, 50/60 Hz
Power consumption approx. 5VA
Operating conditions Temperature: Working temperature range -10 - +40°C
Humidity: 85% RH or less (non-condensing)
External dimensions 44.6 (W) x 110 (H) x 185 (D) [mm] (protruding parts not included)
Weight Approx. 820 g
Accessories
  • AC input cord:
  • 1
  •  
  • Spare fuse(0.5A):
  • 1
  •  
  • Mini-driver:
  • 1
  •  
  • Crimping terminal:
  • 11
  •  
  • Operation Manual:
  • 1

* Nonstandard specifications can support the following: (Please specify when ordering.)
●Current output: 2 -10 mA, Load resistance: 1kΩ or less
●Filter: 10, 100, 1kHz , other
●Calibration value: 0.5mV/V

 
 

Option

Nonstandard specifications are prepared for LC240. Please specify them when ordering.

Configurable parameter Standard specification Changeable specification
Current output 4 - 20mA, Load resistance: 500Ω or less 2 - 10mA, Load resistance: 1kΩ
Filter 1 Hz 10 Hz, 100 Hz, 1kHz
Calibration value 1mV/V 0.5mV/V
 
 

Optional accessories

Model Description
POT03 Potentiometer for external zero adjustment
 
 

Download

Product CatalogueOperation manualExternal dimentionSupport toolsSoftware
GC2011
 
  • User registration is required to download operation manuals, external dimentions, support tools and software.
  • Acrobat Reader software from Adobe Systems is required to view PDF files.